New Shared SEC-MALS column

October 4, 2016

We've added a new shared column for SEC-MALS, an Agilent AdvanceBio 300Å.  

Column: Agilent AdvanceBio 300Å, 7.8 mm x 300 mm

  • Resolution: 5KDa – 1.25 MDa (17 KDa - 667 KDa for MALS)
  • pH range: 2-8.5
  • Max pressure: 150 bar

Experimental conditions

  • Buffer:  25 mM HEPES pH 7.5, 150 mM NaCl, 0.02% NaN3
  • Sample: 100 µl BSA, 2mg/ml (Pierce 23209)
  • Flowrate: 0.5 ml/min
  • Runtime: 35 min
  • Pressure: 67 bar

SEC-MALS Data -BSA